{"id":5474,"date":"2026-08-07T06:55:36","date_gmt":"2026-08-06T21:55:36","guid":{"rendered":"https:\/\/www.aem.imr.tohoku.ac.jp\/?p=5474"},"modified":"2026-08-07T06:55:36","modified_gmt":"2026-08-06T21:55:36","slug":"jem-2100plus-%e8%a9%a6%e6%96%99%e3%82%bb%e3%83%83%e3%83%88%ef%bc%881%e8%bb%b8%e5%82%be%e6%96%9c%e3%83%9b%e3%83%ab%e3%83%80%ef%bc%89","status":"publish","type":"post","link":"https:\/\/www.aem.imr.tohoku.ac.jp\/en\/jem-2100plus-%e8%a9%a6%e6%96%99%e3%82%bb%e3%83%83%e3%83%88%ef%bc%881%e8%bb%b8%e5%82%be%e6%96%9c%e3%83%9b%e3%83%ab%e3%83%80%ef%bc%89\/","title":{"rendered":"JEM-2100plus: Specimen Mounting (Single Tilt Holder)"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\"><em>Note: This page was translated by AI from the Japanese original. In case of any discrepancy, the Japanese version shall prevail.<\/em><\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><em>Original post date (Japanese version): June 22, 2024<\/em><\/p>\n\n\n\n<p class=\"has-background wp-block-paragraph\">&#x1f4a1; Handle the specimen and specimen holder carefully, as they are easily damaged.<\/p>\n\n\n<h3 class=\"wp-block-heading\">Specimen Mounting<\/h3>\n\n\n<ol class=\"wp-block-list\">\n<li>Preparation\n<ol class=\"wp-block-list\">\n<li>Place the specimen holder stand under a digital microscope (or stereomicroscope).<\/li>\n\n\n<li>Gather tools in an easy-to-work position.\n<ol class=\"wp-block-list\">\n<li>Tweezers<\/li>\n\n\n<li>Precision tweezers<\/li>\n\n\n<li>Vacuum tweezers<\/li>\n\n\n<li>Special screwdriver<\/li>\n\n\n<li>Silicone mat (temporary specimen placement)<\/li>\n<\/ol>\n<\/li>\n\n\n<li>Turn on the power of the digital microscope and focus on the specimen exchange stand.<\/li>\n<\/ol>\n<\/li>\n\n\n\n<li>Specimen Preparation\n<ol class=\"wp-block-list\">\n<li>Specimen condition\n<ol class=\"wp-block-list\">\n<li>Must be sufficiently dry. (To avoid deteriorating the vacuum)<\/li>\n\n\n<li>Must be securely fixed. (To prevent damage inside the column, and to prevent drift during observation)<\/li>\n\n\n<li>Magnetization should be small. (To avoid being affected by the magnetic field of the objective lens)<\/li>\n\n\n<li>The area around the observation target should be considered sufficiently thin (100nm or less).<\/li>\n<\/ol>\n<\/li>\n\n\n<li>Take out the specimen and temporarily place it on the silicone mat.\n<ol class=\"wp-block-list\">\n<li>Make it easy to handle when mounting on the holder.\n<ol class=\"wp-block-list\">\n<li>It may be placed on weighing paper, etc.<\/li>\n\n\n<li>You can take it out directly from the grid case. (When using precision tweezers)<\/li>\n<\/ol>\n<\/li>\n<\/ol>\n<\/li>\n<\/ol>\n<\/li>\n\n\n\n<li>Holder Preparation\n<ol class=\"wp-block-list\">\n<li>Take out the holder to be used from the holder station. (Here, the 1-axis holder)\n<ol class=\"wp-block-list\">\n<li>Turn on the power of the holder station<\/li>\n\n\n<li>Valve Open\n<ol class=\"wp-block-list\">\n<li>Wait until it leaks to atmosphere.<\/li>\n<\/ol>\n<\/li>\n\n\n<li>Remove the holder.<\/li>\n<\/ol>\n<\/li>\n\n\n<li>Set the holder on the specimen holder stand. (Place the tip of the holder on the specimen exchange stand)<\/li>\n\n\n<li>Fix the holder with the screws of the holder support.<\/li>\n\n\n<li>Evacuate the holder station\n<ol class=\"wp-block-list\">\n<li>Insert dummy plug\/or holder<\/li>\n\n\n<li>Press the [EVAC] button\n<ol class=\"wp-block-list\">\n<li>Check TMP MONITOR\n<ol class=\"wp-block-list\">\n<li>ACC lights up<\/li>\n\n\n<li>Wait until RUNNING lights up.<\/li>\n<\/ol>\n<\/li>\n<\/ol>\n<\/li>\n\n\n<li>Valve Close<\/li>\n\n\n<li>Turn off the power of the holder station<\/li>\n<\/ol>\n<\/li>\n<\/ol>\n<\/li>\n\n\n\n<li>Specimen Loading\n<ol class=\"wp-block-list\">\n<li>Remove the specimen cartridge (holder tip)\n<ol class=\"wp-block-list\">\n<li>Release the fixation with the cartridge removal tool,<\/li>\n\n\n<li>Remove the cartridge with tweezers,<\/li>\n\n\n<li>Temporarily place it on the silicone mat.<\/li>\n<\/ol>\n<\/li>\n\n\n<li>Loosen the fixing screws with the special screwdriver (3 turns).\n<p class=\"has-background wp-block-paragraph\">&#x1f4a1; Turning it too much will cause the screw to come off.\nIf not loosened enough, the plate holding spring will not move.<\/p>\n<\/li>\n\n\n<li>Rotate the specimen holding plate with precision tweezers.<\/li>\n\n\n<li>Place the specimen on the tip of the holder.\n<ul class=\"wp-block-list\">\n<li>(Observation surface facing down)<\/li>\n<\/ul>\n<\/li>\n\n\n<li>Rotate the specimen holding plate over the specimen with precision tweezers.<\/li>\n\n\n<li>Tighten the fixing screws with the special screwdriver. (2 places)\n<p class=\"has-background wp-block-paragraph\">&#x1f4a1; Do not tighten too strongly. (Just right)\nMust not be loose<\/p>\n<\/li>\n\n\n<li>Attach the specimen cartridge (holder tip)\n<ol class=\"wp-block-list\">\n<li>Release the fixation with the cartridge removal tool,<\/li>\n\n\n<li>Attach the cartridge with tweezers.<\/li>\n<\/ol>\n<\/li>\n<\/ol>\n<\/li>\n<\/ol>\n\n\n<h3 class=\"wp-block-heading\">Specimen Removal<\/h3>\n\n\n<p class=\"has-background wp-block-paragraph\">&#x1f4a1; Reverse procedure of specimen mounting<\/p>\n\n\n<ol class=\"wp-block-list\">\n<li>Pull out the specimen holder from the column<\/li>\n\n\n\n<li>Set the holder on the specimen holder stand. (Place the tip of the holder on the specimen exchange stand)<\/li>\n\n\n\n<li>Fix the holder with the screws of the holder support.<\/li>\n\n\n\n<li>Specimen removal\n<ol class=\"wp-block-list\">\n<li>Remove the specimen cartridge (holder tip)\n<ol class=\"wp-block-list\">\n<li>Temporarily place on the silicone mat<\/li>\n<\/ol>\n<\/li>\n\n\n<li>Loosen the fixing screws with the special screwdriver (3 turns).<\/li>\n\n\n<li>Rotate the specimen holding plate with precision tweezers.<\/li>\n\n\n<li>Remove the specimen.\n<ol class=\"wp-block-list\">\n<li>Return the removed specimen to its case<\/li>\n<\/ol>\n<\/li>\n\n\n<li>Return the specimen holding plate to its original position with precision tweezers.<\/li>\n\n\n<li>Tighten the fixing screws with the special screwdriver. (2 places)<\/li>\n\n\n<li>Attach the specimen cartridge (holder tip)<\/li>\n<\/ol>\n<\/li>\n\n\n\n<li>Return the holder to the holder station.\n<ol class=\"wp-block-list\">\n<li>Turn on the power of the holder station<\/li>\n\n\n<li>Valve Open\n<ol class=\"wp-block-list\">\n<li>Wait until it leaks to atmosphere.<\/li>\n<\/ol>\n<\/li>\n\n\n<li>Remove the dummy plug.<\/li>\n\n\n<li>Insert the holder<\/li>\n\n\n<li>Evacuate the holder station\n<ol class=\"wp-block-list\">\n<li>Insert dummy plug\/or holder<\/li>\n\n\n<li>Press the [EVAC] button\n<ol class=\"wp-block-list\">\n<li>Check TMP MONITOR\n<ol class=\"wp-block-list\">\n<li>ACC lights up<\/li>\n\n\n<li>Wait until RUNNING lights up.<\/li>\n<\/ol>\n<\/li>\n<\/ol>\n<\/li>\n\n\n<li>Valve Close<\/li>\n\n\n<li>Turn off the power of the holder station<\/li>\n<\/ol>\n<\/li>\n<\/ol>\n<\/li>\n<\/ol>\n\n\n\n<p class=\"wp-block-paragraph\"><em>*This post was created by an AI agent.<\/em><\/p>\n","protected":false},"excerpt":{"rendered":"<p>Note: This page was translated by AI from the Japanese original. In case of any discrepancy, the Japanese version shall prevail. Original post date (Japanese version): June 22, 2024 &#x1f4a1; Handle the specimen and specimen holder carefully, as they are easily damaged. Specimen Mounting Specimen Removal &#x1f4a1; Reverse procedure of specimen mounting *This post was <a href=\"https:\/\/www.aem.imr.tohoku.ac.jp\/en\/jem-2100plus-%e8%a9%a6%e6%96%99%e3%82%bb%e3%83%83%e3%83%88%ef%bc%881%e8%bb%b8%e5%82%be%e6%96%9c%e3%83%9b%e3%83%ab%e3%83%80%ef%bc%89\/\" class=\"more-link\">&#8230;<span class=\"screen-reader-text\">  JEM-2100plus: Specimen Mounting (Single Tilt Holder)<\/span><\/a><\/p>\n","protected":false},"author":11,"featured_media":0,"comment_status":"closed","ping_status":"closed","sticky":false,"template":"","format":"standard","meta":{"_locale":"en_US","_original_post":"https:\/\/www.aem.imr.tohoku.ac.jp\/?p=3194","footnotes":"","jetpack_post_was_ever_published":false},"categories":[1],"tags":[],"class_list":["post-5474","post","type-post","status-publish","format-standard","hentry","category-uncategorized","en-US"],"jetpack_featured_media_url":"","_links":{"self":[{"href":"https:\/\/www.aem.imr.tohoku.ac.jp\/wp-json\/wp\/v2\/posts\/5474","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.aem.imr.tohoku.ac.jp\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.aem.imr.tohoku.ac.jp\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.aem.imr.tohoku.ac.jp\/wp-json\/wp\/v2\/users\/11"}],"replies":[{"embeddable":true,"href":"https:\/\/www.aem.imr.tohoku.ac.jp\/wp-json\/wp\/v2\/comments?post=5474"}],"version-history":[{"count":4,"href":"https:\/\/www.aem.imr.tohoku.ac.jp\/wp-json\/wp\/v2\/posts\/5474\/revisions"}],"predecessor-version":[{"id":5768,"href":"https:\/\/www.aem.imr.tohoku.ac.jp\/wp-json\/wp\/v2\/posts\/5474\/revisions\/5768"}],"wp:attachment":[{"href":"https:\/\/www.aem.imr.tohoku.ac.jp\/wp-json\/wp\/v2\/media?parent=5474"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.aem.imr.tohoku.ac.jp\/wp-json\/wp\/v2\/categories?post=5474"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.aem.imr.tohoku.ac.jp\/wp-json\/wp\/v2\/tags?post=5474"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}