💡 本ページはJEM-2100plusのクイックマニュアルです。 内容は変更される場合があります。 実際に装置を使用する場合は、事前にトレーニングを受けて下さい。
- [PANEL-R : Function : MAG1] 像観察モードへ
- [PANEL-L : APERTURE CONTROL :OL(OPEN)] Objective aperture: OPEN
- [HCA] High contrast objective aperture: OPEN
- [SA] Selected field aperture: OPEN
- [PANEL-R : STD FOCUS] Objective lens reference current call
💡 STF FOCUS can be done frequently.
- Probe settings
- [PANEL-L: PROBE CONTROL: TEM] TEM mode
- [PANEL-L : PROBE CONTROL : SPOT SIZE] SPOT SIZE setting (normal: 2)
- [PANEL-L : PROBE CONTROL : αSELECTOR] Convergence angle (irradiation range) setting (normal: 2)
- [CLA] Condenser aperture selection/centering (normal: 2)
- [PANEL-R : MAG/CAM L] Magnification setting(Usually: from about 10 to 50K)
- beam centering
- [PANEL-L: BRIGHTNESS] Focus the beam on the fluorescent screen. (crossover)
- [PANEL-L : DEF/STIG : BRIGHT TILT] 明視野照射モード選択
- [PANEL-L/R: SHIFT-X/Y] Centering with Beam Shift
- Condenser lens astigmatism adjustment
- [PANEL-L: DEF/STIG: COND STIG] Condenser lens astigmatism adjustment mode selection
- [PANEL-L/R: DEF/STIG-X/Y] (Cond STIG) Make the spot shape closer to a perfect circle.
- Brightness adjustment
- [PANEL-L: BRIGHTNESS] Brightness for easy observation: beam spread to the same extent as the screen
- Searching for field of view
- [TB] Move the observation target to the center of the field of view using the trackball
- Focusing (coarse)
- [PANEL-R : STD FOCUS] 基準電流呼び出し
- [PANEL-R: Z△▽] Sample height (Z) adjustment: Lowest contrast
- Objective lens astigmatism correction
- Medium double
- Observing the vicinity of the membrane hole (x20K ~)
Center the beam and adjust the brightness for easy observation.
- [PANEL-L: DEF/STIG: OBJ STIG] Objective lens astigmatism adjustment mode selection
- [PANEL-R: OBJ FOCUS] Display Fresnel fringes with hyperfocus
- [PANEL-L/R: DEF/STIG-X/Y] Adjust the Fresnel fringe to be isotropic
- Observing the vicinity of the membrane hole (x20K ~)
- High magnification
- 非晶質領域(支持膜など)を観察(x100K~)
Center the beam and adjust the brightness for easy observation.
- [PANEL-L: DEF/STIG: OBJ STIG] Objective lens astigmatism adjustment mode selection
- [PANEL-R : OBJ FOCUS] &[PANEL-L/R: DEF/STIG-X/Y]
- Confirm that the image flows due to hyperfocal <> underfocal
- Adjust to minimize image flow
- 非晶質領域(支持膜など)を観察(x100K~)
- Medium double
- Insert objective aperture (if necessary)
- (Simple)
- [OLA/HCA] Objective aperture (in-gap orhigh contrast) insert
- [OLA/HCA] Centering of objective aperture
- (推奨)
- [SA] Inserting and centering the selected area diaphragm
- [PANEL-R : Function : DIFF] To diffraction mode: Electron diffraction display
- [OLA/HCA] Objective aperture (in gapor high contrast) and select transmitted wave.
- [PANEL-R : Function : MAG1] Return to image observation mode
- [SA] Pull out the selected area diaphragm
- (Simple)
- Focusing (precision)
- [PANEL-R: OBJ FOCUS] Adjust to a focus that is easy to see near the positive focus
- 正焦点(ジャストフォーカス):膜孔の輪郭のコントラスト最小
- Hyperfocal (overfocus): Focus clockwise from positive focus
- Underfocus: Move the focus from positive focus to counterclockwise.
💡 Samples with low contrast will be unclear at positive focus, so set the focus to be slightly underfocused.
- [PANEL-R: OBJ FOCUS] Adjust to a focus that is easy to see near the positive focus
- (photograph)
- [Screen] Lower the magnification for CCD camera photography.
- [Screen] Adjust brightness
- Display the image with a CCD camera and adjust focus and brightness
- take a photo