💡 This page is a quick manual for JEM-2100plus. Contents are subject to change. Please receive training before actually using the equipment.
Observation window (large fluorescent screen)
- Remove the light-blocking lid.
- Observe with a fluorescent screen.
- [TC: TEM System Task Bar: Screen] Brightness check (if necessary)
- Beam current density on large fluorescent screen
- Attach the blackout lid.
- Always cover the container when handling liquid nitrogen.
- Preventing observation window glass from breaking
- Close the lid when using the CCD camera/STEM.
- Preventing noise from shooting data
- Always cover the container when handling liquid nitrogen.
Small fluorescent screen/binoculars
💡 If precise observation or focusing is required, binocular magnifiers and a small fluorescent screen can be used (magnification factor x10).
- Push down the small fluorescent plate lever on the right side of the lens barrel.
- A small fluorescent screen is inserted into the electron beam path.
- Move the binoculars in front of the fluorescent screen.
- Focus the binoculars on the small fluorescent screen.
- While looking through the binoculars, move the focusing knob to focus on the center point of the small fluorescent screen.
- If the left and right focal points are different, adjust by rotating the eyepiece.
- [TC: TEM System Task Bar: Screen] Brightness check (if necessary)
- Displays beam current density on small fluorescent screen
beam stopper
💡 Can be used to block transmitted waves in electron diffraction or as a landmark during image observation.
- Turn the beam stopper lever on the left side of the lens barrel.
- A beam stopper is inserted into the electron beam path.
- If necessary, turn the lever slightly to fine-tune the position.
- If the stopper movement is too large and fine adjustment is difficult, move the beam.
screen operation
💡 Open and close the large fluorescent screen when using a bottom mount camera (CCD camera).
- An enlarged image or electron diffraction pattern is displayed on a fluorescent screen.
- Lower the magnification and camera length.
- The field of view of the camera is narrower than that of the screen.
- Adjust brightness (darken)
- Lowering the magnification increases the current density and may damage the sensor.
- Insert beam stopper (if required)
- [PANEL R: F4] Raise the screen (90 degrees)
- [PANEL R: F4] Lower the screen (0 degrees)
💡 Can also be operated from the TEM center