💡 Display the beam on the screen. If it is not displayed, remove the cause.
- Prerequisite
- Routine adjustments have been made by the administrator.
- Preliminary check
- The specimen holder is set
- Accelerating voltage (usually 200kV) is applied.
- Electron beam generation (Reference) JEM-2100plus Electron Beam Generation/Stopping (Basics)
- Check the vacuum level
- SIP 10^-5Pa range
- Electron beam generation
- Beam Current About 103~107uA
- If Beam Current is small, BIAS adjustment is required (contact the administrator)
- If Beam Current does not change, there is a possibility of filament breakage (contact the administrator)
- Check the vacuum level
- Change to Image Mode
- [PANEL-L : Function : MAG1(MAG2)]
- Confirm the beam on the screen
- If the beam is displayed on the screen, it is OK (work completed)
- If the beam is not displayed on the screen
- Remove all apertures from the electron beam path
- Possibility that an aperture is blocking the electron beam path
- Lower the magnification
- Possibility that the beam is reaching outside the field of view
- Check the probe
- Possibility that the beam is too dark
- Move specimen
- Possibility that the specimen is blocking the electron beam path
- Remove all apertures from the electron beam path
- If the beam is not displayed on the screen even after performing all the above operations, alignment is required.
- Contact the administrator
- You may try loading an alignment file.
- Be sure to load the latest version