Overview
- Confirmation of functions of TEM Center, etc.
- Confirmation of functions and operation methods for each element
- Observation on the screen
- Observation using the camera will be in the next session
Program
- Name/Function Confirmation
- TEM Center
- Physical controllers
- Other operations
- Specimen Shift
- Specimen shift using trackball (X/Y)
- Confirmation of specimen set orientation and screen orientation
- (Confirm image rotation due to MAG/LowMag/magnification change)
- Confirm that the limit works when moved significantly
- Specimen tilt (TILT-X/Y)
- Confirm changes in appearance when tilted significantly
- Confirm that the limit works when moved significantly
- Specimen fine movement using Piezo drive (SFX/SFY)
- Specimen height adjustment (Z)
- ImageX/Y
- Stage initialization
- Stage memory and recall
- Specimen shift using trackball (X/Y)
- Aperture Operation
- Selected area aperture
- On the optical axis in image mode
- High contrast objective aperture
- On the optical axis in image mode
- (Recommended) On the optical axis in DIFF mode
- In-gap objective aperture
- On the optical axis in image mode
- (Recommended) On the optical axis in DIFF mode
- Condenser lens aperture
- On the optical axis in image mode
- Cannot be opened
- (Reference) Illumination system adjustment
- On the optical axis in image mode
- Hard X-ray aperture
- Selected area aperture
- Imaging System Selection
- Imaging mode (TEM/DIFF) switching
- TEM Adjustment
- Magnification change
- Details -> Image Observation (Basics)
- DIFF Adjustment
- Camera length change
- Details -> Electron Diffraction (Basics)
- Illumination System Adjustment
- Brightness Adjustment
- Confirm that converging the beam makes it brighter, expanding makes it darker.
- (Simple Centering) BeamAlign
- Center the converged beam on the screen
- Condenser lens astigmatism correction
- SPOT Size / α Selection
- Condenser Lens Alignment GunAlign/BeamAlign
- GunTilt Adjustment (Simple)
- Condenser aperture centering
- Brightness Adjustment
- Objective Lens Astigmatism Correction
- Voltage/Current Center Alignment
- Current axis alignment
- Voltage axis alignment
- Beam Display
- Display the beam on the screen. If the beam is not displayed even when Emission is On, check the following.
- Beam is generated -> Beam Current
- Not blocked by an aperture -> Remove objective/selected area apertures
- Not blocked by the specimen -> Move specimen / empty holder
- Appropriate magnification and Brightness (Curr. Density)
- Axis alignment of illumination system
- Display the beam on the screen. If the beam is not displayed even when Emission is On, check the following.
- Screen
- Brightness adjustment on the screen
- Appearance and CurrDensity
- (MagLink)
- Binoculars
- Screen open/close
- Brightness adjustment on the screen
- Image Observation (Basics)
- Training uses an Al-deposited specimen
- Electron Diffraction (Basics)
- Training uses an Al-deposited specimen
Axis Alignment performed in Beginner Level
Voltage axis adjustmentCurrent axis adjustment- Illumination system interlocking ratio adjustment
- Tilt adjustment (Tilt 2-stage deflection adjustment)
- Horizontal adjustment (Shift 2-stage deflection adjustment)