💡 This page is a quick manual for the JEM-2100plus. The contents are subject to change. Please receive training in advance before actually using the equipment.
💡 Basic axis alignments are assumed to have been completed by the administrator.
BRIGHTNESS Adjustment
- [PANEL-L : BRIGHTNESS] Brightness adjustment
- Turn it fully clockwise once (maximum excitation/alarm sounds), then return it counterclockwise (converge).
- Clockwise: CL3 excitation strengthens: Focal length shortens
- When turned fully, the beam diverges (becomes darker).
- Counterclockwise: CL3 excitation weakens: Focal length lengthens
- When turned fully, the beam diverges (becomes darker)
- Clockwise: CL3 excitation strengthens: Focal length shortens
- Turn it fully clockwise once (maximum excitation/alarm sounds), then return it counterclockwise (converge).
- Brightness: Coarse/Fine adjustment selection
- [PANEL-L : BRIGHTNESS: CRS] While selected, the backlight turns on and the changing speed becomes x16.
- Brightness confirmation
- Confirm visually on the screen
- [TC: TEM System Task Bar : SCREEN]
Beam Centering
- [PANEL-R : STD FOCUS] Recall objective lens reference current
- Just in case
- [PANEL-L : BRIGHTNESS] Converge the beam on the fluorescent screen.
- [PANEL-L : DEF/STIG : BRIGHT TILT] Select bright field illumination mode
- [PANEL-L/R: SHIFT-X/Y] Move the beam to the center of the fluorescent screen with Beam Shift
Condenser Lens Astigmatism Correction
- [PANEL-L : BRIGHTNESS] Converge the beam on the fluorescent screen.
- [PANEL-L : DEF/STIG : COND STIG] Select condenser lens astigmatism adjustment mode
- [PANEL-L/R: DEF/STIG-X/Y] (Cond STIG) Make the spot shape closer to a perfect circle.
SPOT SIZE / α Selection (TEM Mode)
- SPOT SIZE: Size of the electron beam when most converged
- Smaller is brighter
- α: Convergence angle when the electron beam is focused on the specimen. (Size of the parallel illumination area)
- [PANEL-R : STD FOCUS] Recall objective lens reference current
- Just in case
- [PANEL-L : PROBE CONTROL : SPOT SIZE] SPOT SIZE setting (1 to 5: Normal: 2)
- [PANEL-L : PROBE CONTROL : αSELECTOR] Convergence angle (illumination area) setting (1 to 9: Normal: 2)
- Perform Beam centering and Condenser lens astigmatism correction (if necessary)
Gun Tilt Adjustment (Simple)
💡 Make the electron beam emitted from the GUN incident straight onto the specimen along the optical axis. Assuming other optical systems are well adjusted.
- [PANEL-R : Function : MAG1] To image observation mode
- [PANEL-L : APERTURE CONTROL : OL(OPEN)] Objective aperture: OPEN
- [HCA] High contrast objective aperture: OPEN
- [SA] Selected area aperture: OPEN
- [PANEL-R : MAG/CAM L] Magnification setting (Normal: around x5000)
- [PANEL-R : STD FOCUS] Recall objective lens reference current
- Just in case
- Beam centering
- [PANEL-L : BRIGHTNESS] Converge the beam on the fluorescent screen.
- [PANEL-L : DEF/STIG : BRIGHT TILT] Select bright field illumination mode
- [PANEL-L/R: SHIFT-X/Y] Center with Beam Shift
- GUN TILT adjustment
- [TEM Center: Upper menu] Control → Alignment Panel (If not displayed)
- Alignment Panel (for JEM Administrators) will be displayed.
- [TC : Alignment Panel : Deflector : GUN Align] Select GUN alignment mode
- [PANEL-L/R: DEF/STIG-X/Y] Make the beam brightest with Beam Tilt.
- [TEM Center: Upper menu] Control → Alignment Panel (If not displayed)
Condenser Lens Alignment
💡 Reduce beam escape when changing SPOT SIZE.
- Adjust GUN Tilt in advance. (If necessary)
- [PANEL-R : Function : MAG1] To image observation mode
- [PANEL-L : APERTURE CONTROL : OL(OPEN)] Objective aperture: OPEN
- [HCA] High contrast objective aperture: OPEN
- [SA] Selected area aperture: OPEN
- [PANEL-R : MAG/CAM L] Magnification setting (Normal: around 50K)
- [PANEL-R : STD FOCUS] Recall objective lens reference current
- Just in case
- [PANEL-L : BRIGHTNESS] Converge the beam on the fluorescent screen.
- Adjust at SPOT SIZE 1
- [PANEL-L : PROBE CONTROL : SPOT SIZE] Select 1
- [TEM Center: Upper menu] Control → Alignment Panel (If not displayed)
- Alignment Panel (for JEM Administrators) will be displayed.
- [TC : Alignment Panel : Deflector : GUN Align] Select GUN alignment mode
- [PANEL-L/R: SHIFT-X/Y] Move the beam to the center of the fluorescent screen with Beam Shift
- Adjust at SPOT SIZE 5
- [PANEL-L : PROBE CONTROL : SPOT SIZE] Select 5
- [PANEL-L : DEF/STIG : BRIGHT TILT] or [TC : Alignment Panel : Deflector : Bright Tilt ] Select bright field illumination mode
- [PANEL-L/R: SHIFT-X/Y] Move the beam to the center of the fluorescent screen with Beam Shift
- Repeat [Adjust at SPOT SIZE 1] and [Adjust at SPOT SIZE 5]
Condenser Aperture Centering
(Reference) JEM-2100plus: Aperture Operation: Condenser Aperture
💡 If the beam does not change concentrically (swings) when BRIGHTNESS is changed, the condenser aperture is misaligned from the optical axis, so adjust referring to the above.