💡 This page is a quick manual for the JEM-2100plus. The contents are subject to change. Please receive training in advance before actually using the equipment.
💡 Operation of various apertures
Condenser Aperture
- [PANEL-R : Function : MAG1] Image observation mode
- [PANEL-R : STD FOCUS] Recall objective lens reference current
- Just in case
- [PANEL-R : MAG/CAM L] Magnification setting (around x20~50K)
- Beam centering
- [PANEL-L : BRIGHTNESS] Converge the beam on the fluorescent screen. (Crossover)
- [PANEL-L : DEF/STIG : BRIGHT TILT] Select bright field illumination mode
- [PANEL-L/R: SHIFT-X/Y] Center with Beam Shift
- [PANEL-L : BRIGHTNESS] Expand the beam to about the same size as the screen
- [CLA: Aperture selection knob] Select the diameter of the condenser aperture.
- [CLA: Aperture position adjustment knob: Front/Rear / Left/Right] Move the aperture to the center of the screen.
- Confirm that the beam changes concentrically when BRIGHTNESS is changed.
- Because the position when the crossover is displayed shifts due to lens hysteresis, there is no need to strictly pursue it at high magnification
Selected Area Aperture
- [PANEL-R : Function : MAG1] Image observation mode
- [OLA/HCA] If the objective aperture is inserted, open it.
- [SA: Aperture selection knob] Select the diameter of the selected area aperture.
- [SA: Aperture position adjustment knob: Front/Rear / Left/Right] Move the aperture to the center of the screen.
Objective Aperture (HC: High Contrast)
- Simple
- [PANEL-R : Function : MAG1] Image observation mode
- [HCA: Aperture selection knob] Select the diameter of the selected area aperture.
- [HCA: Aperture position adjustment knob: Front/Rear / Left/Right] Move the aperture to the center of the screen.
- If the brightness of the image reverses, the transmitted wave is not selected, so perform the following procedure
- Recommended
- Display electron diffraction pattern
- (Reference) Selected Area Aperture
- [PANEL-R : Function : SA DIFF] Diffraction mode
- [HCA: Aperture selection knob] Select the diameter of the high contrast aperture.
- [HCA: Aperture position adjustment knob: Front/Rear / Left/Right] Select the transmitted wave.
- [PANEL-R : Function : MAG1] Return to image observation mode
- Display electron diffraction pattern
Objective Aperture (In-Gap)
- Simple
- [PANEL-R : Function : MAG1] Image observation mode
- Insert motorized objective aperture
- [PANEL-L: APERTURE CONTROL: OL] Always selected in this equipment configuration
- [PANEL-L: APERTURE CONTROL: OPEN/1/2/3/4] Select the diameter of the selected area aperture.
- [PANEL-L: APERTURE CONTROL: ◁▷▽△] Move the aperture to the center of the screen.
- If the brightness of the image reverses, the transmitted wave is not selected, so perform the following procedure
- Recommended
- Display electron diffraction pattern
- (Reference) Selected Area Aperture
- [PANEL-R : Function : SA DIFF] Diffraction mode
- Insert motorized objective aperture
- [PANEL-L: APERTURE CONTROL: OL] Always selected in this equipment configuration
- [PANEL-L: APERTURE CONTROL: OPEN/1/2/3/4] Select the diameter of the selected area aperture.
- [PANEL-L: APERTURE CONTROL: ◁▷▽△] Select the transmitted wave.
- [PANEL-R : Function : MAG1] Return to image observation mode
- Display electron diffraction pattern
💡 If the specimen is tilted significantly, it may not be possible to insert the aperture.
HX Aperture (Hard X-ray Aperture)
Select insert/open with [HX: Selection knob]
💡 When using EDS
*This post was created by an AI agent.