STEM-BF
- Display a normal image in TEM mode.
- Objective aperture and selected field aperture are open.
- Switch to SCAN(STEM) mode
- [TC: TEM System Task Bar : Illumination System ] Click the [SCAN] button
- Insert the bright field image detector (on the right side of the lens barrel) into the lens barrel.
- Cover the observation window.
- Operation image display
- [TC: TEM System Task Bar : Image Forming System: Image Selector : BF ] BF detector selection
- [TC: TEM System Task Bar : Image Forming System: Camera Length ] Camera length setting
- Increasing the camera length increases contrast
- [TC:STEM Image Viewer : Preset Mode : Search Mode] スキャン速度選択(高速)
- Brightness/contrast adjustment
- (If necessary) [TC: Top menu: View: Histgram] Histogram display
- [TC: STEM Image Viewer : Contrast ] Contrast adjustment
- [TC: STEM Image Viewer : Brightness ] Brightness adjustment
💡 Automatic adjustment
- height adjustment
- [PANEL-R: STD FOCUS] Reference current call
- [PANEL-R: Z△▽] Sample height (Z) adjustment: Obtain the sharpest image.
- If it is difficult to match, lower the magnification.
- Focusing/stigma correction
- [PANEL-R:OBJ Focus] Focusing
- [PANEL-L: DEF/STIG: COND STIG] Focusing lens astigmatism adjustment mode selection
- [PANEL-L/R: DEF/STIG-X/Y] Adjust to minimize image flow
- [TC:STEM Image Viewer : Preset Mode : Focus Mode] Scan speed selection (for focusing)
- Repeat a-c to get the sharpest image
- photograph
- [TC:STEM Image Viewer : Preset Mode : Photo Mode] Scan speed selection (for shooting: low speed)
- [TC:STEM Image Viewer ] Capture Single
- Exit (switch to TEM mode)
- [TC: TEM System Task Bar : Illumination System ] Click the [TEM] button
- pull out the detector
STEM-DF
- Display a normal image in TEM mode.
- Objective aperture and selected field aperture are open.
- Switch to SCAN(STEM) mode
- [TC: TEM System Task Bar : Illumination System] Click the [SCAN] button
- [TC: TEM System Task Bar : Image Forming System: Image Selector : DF ] DF detector selection
- [TC: TEM System Task Bar : Image Forming System: Camera Length ] Camera length setting (10 or 20cm)
- Ronchigram adjustment
- screen work
- [TC: Image Forming System: Ronchigram Mode: On ] Ronchigram mode On (SCAN stop)
- [CLA#1(φ150um)] Insert centering
💡 Align the center of the aperture with the Ronchigram center
- camera work
- [GMS : TEM Imaging : View] (Exposure : 0.1sec) カメラ挿入・表示
- [PANEL R : F4] Screen open
- Ronchigram Centering
- [PANEL-L: DEF/STIG: PLA] Projection lens alignment mode selection
- [PANEL-L/R: DEF/STIG-X/Y] Centering the transmitted wave to the center of the fluorescent screen
💡 Align the center of Ronchigram with the center of the screen.
- [CLA#3 (φ50um)] Insertion centering
💡 Align the center of Ronchigram and the center of CLA with the center of the screen.
- [TC: Image Forming System: Ronchigram Mode : Off ] Ronchigram mode On (SCAN start)
- [GMS : TEM Imaging : View] Camera stop
- screen work
- Cover the observation window.
- Insert the dark field image detector (left side of the lens barrel) into the lens barrel.
- Operation image display
- [TC: TEM System Task Bar : Image Forming System: Camera Length ] Camera length setting
- [TC:STEM Image Viewer : Preset Mode : Search Mode] Scan speed selection (high speed)
- Brightness/contrast adjustment
- (If necessary) [TC: Top menu: View: Histgram] Histogram display
- [TC: STEM Image Viewer : Contrast ] Contrast adjustment
- [TC: STEM Image Viewer : Brightness ] Brightness adjustment
💡 Automatic adjustment
- height adjustment
- [PANEL-R: STD FOCUS] Reference current call
- [PANEL-R: Z△▽] Sample height (Z) adjustment: Obtain the sharpest image.
- If it is difficult to match, lower the magnification.
- Focusing/stigma correction
- [PANEL-R:OBJ Focus] Focusing
- [PANEL-L: DEF/STIG: COND STIG] Focusing lens astigmatism adjustment mode selection
- [PANEL-L/R: DEF/STIG-X/Y] Adjust to minimize image flow
- [TC:STEM Image Viewer : Preset Mode : Focus Mode] Scan speed selection (for focusing)
- Repeat a-c to get the sharpest image
- photograph
- [TC:STEM Image Viewer : Preset Mode : Photo Mode] Scan speed selection (for shooting: low speed)
- [TC:STEM Image Viewer ] Capture Single
- Exit (switch to TEM mode)
- [TC: TEM System Task Bar : Illumination System ] [TEM]ボタンクリック
- pull out the detector