💡 Handle the specimen and specimen holder carefully, as they are easily damaged.
Specimen Mounting
- Preparation
- Place the specimen holder stand under a digital microscope (or stereomicroscope).
- Gather tools in an easy-to-work position.
- Tweezers
- Precision tweezers
- Vacuum tweezers
- Special screwdriver
- Silicone mat (temporary specimen placement)
- Turn on the power of the digital microscope and focus on the specimen exchange stand.
- Specimen Preparation
- Specimen condition
- Must be sufficiently dry. (To avoid deteriorating the vacuum)
- Must be securely fixed. (To prevent damage inside the column, and to prevent drift during observation)
- Magnetization should be small. (To avoid being affected by the magnetic field of the objective lens)
- The area around the observation target should be considered sufficiently thin (100nm or less).
- Take out the specimen and temporarily place it on the silicone mat.
- Make it easy to handle when mounting on the holder.
- It may be placed on weighing paper, etc.
- You can take it out directly from the grid case. (When using precision tweezers)
- Make it easy to handle when mounting on the holder.
- Specimen condition
- Holder Preparation
- Take out the holder to be used from the holder station. (Here, the 1-axis holder)
- Turn on the power of the holder station
- Valve Open
- Wait until it leaks to atmosphere.
- Remove the holder.
- Set the holder on the specimen holder stand. (Place the tip of the holder on the specimen exchange stand)
- Fix the holder with the screws of the holder support.
- Evacuate the holder station
- Insert dummy plug/or holder
- Press the [EVAC] button
- Check TMP MONITOR
- ACC lights up
- Wait until RUNNING lights up.
- Check TMP MONITOR
- Valve Close
- Turn off the power of the holder station
- Take out the holder to be used from the holder station. (Here, the 1-axis holder)
- Specimen Loading
- Remove the specimen cartridge (holder tip)
- Release the fixation with the cartridge removal tool,
- Remove the cartridge with tweezers,
- Temporarily place it on the silicone mat.
- Loosen the fixing screws with the special screwdriver (3 turns).
💡 Turning it too much will cause the screw to come off. If not loosened enough, the plate holding spring will not move.
- Rotate the specimen holding plate with precision tweezers.
- Place the specimen on the tip of the holder.
- (Observation surface facing down)
- Rotate the specimen holding plate over the specimen with precision tweezers.
- Tighten the fixing screws with the special screwdriver. (2 places)
💡 Do not tighten too strongly. (Just right) Must not be loose
- Attach the specimen cartridge (holder tip)
- Release the fixation with the cartridge removal tool,
- Attach the cartridge with tweezers.
- Remove the specimen cartridge (holder tip)
Specimen Removal
💡 Reverse procedure of specimen mounting
- Pull out the specimen holder from the column
- Set the holder on the specimen holder stand. (Place the tip of the holder on the specimen exchange stand)
- Fix the holder with the screws of the holder support.
- Specimen removal
- Remove the specimen cartridge (holder tip)
- Temporarily place on the silicone mat
- Loosen the fixing screws with the special screwdriver (3 turns).
- Rotate the specimen holding plate with precision tweezers.
- Remove the specimen.
- Return the removed specimen to its case
- Return the specimen holding plate to its original position with precision tweezers.
- Tighten the fixing screws with the special screwdriver. (2 places)
- Attach the specimen cartridge (holder tip)
- Remove the specimen cartridge (holder tip)
- Return the holder to the holder station.
- Turn on the power of the holder station
- Valve Open
- Wait until it leaks to atmosphere.
- Remove the dummy plug.
- Insert the holder
- Evacuate the holder station
- Insert dummy plug/or holder
- Press the [EVAC] button
- Check TMP MONITOR
- ACC lights up
- Wait until RUNNING lights up.
- Check TMP MONITOR
- Valve Close
- Turn off the power of the holder station