💡 Specimens and specimen holders are fragile, so handle them with care.
Specimen Mounting
- Preparation
- Place the specimen holder stand under a digital microscope (or stereomicroscope).
- Gather tools in an easy-to-work position.
- Precision tweezers
- Vacuum tweezers
- Dedicated screwdriver
- Silicone mat (temporary specimen placement)
- Turn on the digital microscope and focus on the specimen exchange stand.
- Specimen Preparation
- Specimen condition
- Must be sufficiently dry. (To prevent vacuum degradation)
- Must be securely fixed. (To prevent damage in the column and drift during observation)
- Must have low magnetization (to avoid being affected by the objective lens magnetic field)
- The area around the observation target must be sufficiently thin (considered to be 100nm or less).
- Take out the specimen and temporarily place it on the silicone mat.
- Make it easy to handle when loading the holder.
- You may place it on powder paper, etc.
- You may take it out directly from the grid case. (When using precision tweezers)
- Make it easy to handle when loading the holder.
- Specimen condition
- Holder Preparation
- Take out the holder to be used from the holder station. (Here, a double-tilt beryllium holder)
- Turn on the holder station power
- Open the valve
- Wait until it leaks to the atmosphere.
- Take out the holder.
- Confirm that the Y-axis tilt of the stage is 0. (Check by looking directly at the holder)
💡 If the stage is tilted, do not place it on the holder stand. Risk of holder damage.
- Set the holder on the specimen holder stand. (Place the tip of the holder on the specimen exchange stand)
- Fix the holder with the holder support screw.
- Evacuate the holder station
- Insert a dummy plug/or holder
- Press the [EVAC] button
- Check TMP MONITOR
- ACC lights up
- Wait until RUNNING lights up.
- Check TMP MONITOR
- Close the valve
- Turn off the holder station power
- Take out the holder to be used from the holder station. (Here, a double-tilt beryllium holder)
- Specimen Loading
- Loosen the fixing screw with the dedicated screwdriver (3 turns).
💡 If you turn it too much, the screw will come off. If not loosened enough, the plate holding spring will not move.
- Move the plate holding spring off the specimen fixing plate with precision tweezers.
- Remove the specimen fixing plate and spacer, and temporarily place them on the silicone mat.
- (Use precision tweezers and vacuum tweezers)
- Use vacuum tweezers to place the specimen on the tip of the holder.
- (Observation surface facing down)
- Place the spacer on top of the specimen.
- (Not required if the specimen thickness is 100um or more.)
- Place the specimen fixing plate on top of the spacer.
- Move the plate holding spring onto the specimen fixing plate. (2 locations)
- Tighten the fixing screw with the dedicated screwdriver. (2 locations)
💡 Do not tighten too strongly. (Just the right strength) Make sure it is not loose
- Loosen the fixing screw with the dedicated screwdriver (3 turns).
Specimen Removal
💡 Reverse procedure of specimen mounting
- Pull out the specimen holder from the column
- Set the holder on the specimen holder stand. (Place the tip of the holder on the specimen exchange stand)
- Fix the holder with the holder support screw.
- Specimen Removal
- Loosen the fixing screw with the dedicated screwdriver (3 turns).
- Move the plate holding spring off the specimen fixing plate with precision tweezers.
- Remove the specimen fixing plate and spacer, and temporarily place them on the silicone mat.
- Remove the specimen from the tip of the holder and place it on the silicone mat.
- You may return it directly to the grid case.
- Place the spacer on the tip of the holder.
- Place the specimen fixing plate on top of the spacer.
- Move the plate holding spring onto the specimen fixing plate. (2 locations)
- Tighten the fixing screw with the dedicated screwdriver. (2 locations)
- Return the holder to the holder station.
- Turn on the holder station power
- Open the valve
- Wait until it leaks to the atmosphere.
- Remove the dummy plug.
- Insert the holder
- Evacuate the holder station
- Insert a dummy plug/or holder
- Press the [EVAC] button
- Check TMP MONITOR
- ACC lights up
- Wait until RUNNING lights up.
- Check TMP MONITOR
- Close the valve
- Turn off the holder station power