- Pre-insertion Check (TEM Main Body)
- Specimen holder is not inserted
- [TEM Center: Upper menu] Control → Beam
- Display Beam Controller (for JEM Administrators)
- Check accelerating voltage and emission
- [TEM Center: Upper menu] View → Valve/Vacuum Monitor
- Display Valve/Vacuum Monitor
- Check vacuum level
- [TEM Center: Upper menu] View → Condition Monitor
- Display Condition Monitor
- Check alarms
- [TEM Center: Upper menu] Control → Aperture
- Display Aperture Panel Controller
- Retract the in-gap objective lens aperture
- Select [OPEN] in Hole Diameter Select
- [TEM Center: Left side TEM System Task Bar] Stage
- Check stage position and tilt
- Pre-insertion Check (Specimen/Holder)
- Specimen check
- The specimen is correctly fixed and does not move.
- No adhesion of sebum, dust, etc., near the specimen.
- Ion cleaning has been performed if necessary.
- Holder check
- The tip of the holder is not damaged or deformed.
- No dust is attached to the O-ring. No scratches.
- (For 2-axis holder) Y tilt is 0.
- Check both the value on the TEM center and the actual stage tilt
- Specimen check
- Holder Insertion
- Hold the holder handle and the front of the O-ring with both hands, align the guide pin, and insert it straight into the goniometer.
- Push it in until it stops, keeping your hand on it.
- At this position, the insertion detection switch operates. (It will be released if rotated, so do not move your hand)
- Goniometer: Switch the PUMP/AIR switch to PUMP.
- The specimen pre-chamber is evacuated.
- Wait until the green lamp lights up and PiG4 becomes EvacReady (green).
- Insert the holder
- Turn the holder clockwise (slightly). (Until it stops)
- The holder is drawn in. (Slightly)
- Turn the holder clockwise (largely). (Until it stops)
- The holder is (largely) drawn in. (Do not let go until it stops)
- (For 2-axis holder) Connect the cable for Y-axis tilt.
💡 In the above work
- Select the holder
- When the holder is inserted, the Specimen pane is displayed.
- [TEM Center: Upper menu] Control → Specimen
- Select the inserted holder from the Specimen Holder drop-down list.
- Click the [OK] button
- Since the movement and tilt range of the stage differs for each holder, be sure to select the inserted holder.
- When the holder is inserted, the Specimen pane is displayed.
- Wait for the vacuum to recover
- Penning Gauge 20 or less
- SIP 10^-5Pa range
- Hold the holder handle and the front of the O-ring with both hands, align the guide pin, and insert it straight into the goniometer.
Manual Video (youtube channel)