💡 This page is a quick manual for the JEM-2100plus. The contents are subject to change. Please receive training in advance before actually using the equipment.
💡 Specimen Shift and Tilt
Specimen Shift (Motor Driven) (X/Y)
- [TC: TEM System Task Bar : Stage: X/Y] Check specimen position
- [TC: TEM System Task Bar : Stage] Select movement speed
- Becomes faster in the order of ▶⇒▶▶⇒▶▶▶
- Move specimen with [TB: Trackball]
- Move independently for each axis with [TB : SHIFT +X/-X/+Y/-Y] buttons
- Coarse/Fine (Movement Speed) Switching
- Switch with [TB: CRS]. When lit: Coarse
- [TB: CRS] is common to specimen shift (SHIFT-X/Y/Z:TILT-X/Y) and specimen tilt
- Initialize Specimen Position (X/Y)
- [TEM Center: Upper menu] Control → STAGE
- Display Stage Controller
- Select [XY Neutral]
- Confirm with [TC: TEM System Task Ba : Stager]
Specimen Shift (Piezo Driven) (SFX/SFY)
- [TC: TEM System Task Bar : Stage: SFX/SFY] Check specimen position
- Relative position based on X/Y. ~±100% (0.2um)
- [TB: SUPER FINE] Piezo drive mode switching: On when lit
- [TC: TEM System Task Bar : Stage] Select movement speed
- Becomes faster in the order of ▶⇒▶▶⇒▶▶▶
- Move specimen with [TB: Trackball] or [TB : SHIFT +X/-X/+Y/-Y]
- Coarse/Fine (Movement Speed) Switching
- Switch with [TB: CRS]. When lit: Coarse
- [TB: CRS] is common to specimen shift (SHIFT-X/Y/Z:TILT-X/Y) and specimen tilt
- Initialize Specimen Position (SFX/SFY)
- [TEM Center: Upper menu] Control → STAGE
- Display Stage Controller
- Select [SuperFIine Neutral]
- Confirm with [TC: TEM System Task Ba : Stager]
Specimen Tilt (TX/TY)
- [TC: TEM System Task Bar : Stage: TX/TY] Check specimen tilt
- [TC: TEM System Task Bar : Stage] Select movement speed
- Becomes faster in the order of ▶⇒▶▶⇒▶▶▶
- Tilt specimen with [SC : TILT +X/-X/+Y/-Y] buttons
- Coarse/Fine (Movement Speed) Switching
- Switch with [TB: CRS]. When lit: Coarse
- [TB: CRS] is common to specimen shift (SHIFT-X/Y/Z:TILT-X/Y) and specimen tilt
- Initialize Specimen Tilt (TX/TY)
- [TEM Center: Upper menu] Control → STAGE
- Display Stage Controller
- Select [TX/TY Neutral]
- Initialize for each axis
- Confirm with [TC: TEM System Task Ba : Stager]
Specimen Shift (Z)
- [TC: TEM System Task Bar : Stage: Z] Check specimen height
- [TC: TEM System Task Bar : Stage] Select movement speed
- Becomes faster in the order of ▶⇒▶▶⇒▶▶▶
- Move specimen height with [SC :△Z/▽Z] buttons
- Coarse/Fine (Movement Speed) Switching
- Switch with [TB: CRS]. When lit: Coarse
- [TB: CRS] is common to specimen shift (SHIFT-X/Y/Z:TILT-X/Y) and specimen tilt
- Initialize Specimen Tilt (Z)
- [TEM Center: Upper menu] Control → STAGE
- Display Stage Controller
- Select [Z Neutral]
- Confirm with [TC: TEM System Task Ba : Stager]
Stage Initialization
- [TEM Center: Upper menu] Control → STAGE
- Display Stage Controller
- Select [Stage Neutral]
- Confirm with [TC: TEM System Task Ba : Stager]
Stage Position Memory/Recall
- Specimen Position Memory
- [TEM Center: Upper menu] Control → Nanospace Map
- Display Nanospace Map
- Click [Memory] button
- The current holder position is memorized in the specimen position memory list.
- SFX/SFY are not memorized
- Add a comment in the Memo field if necessary
- [TEM Center: Upper menu] Control → Nanospace Map
- Specimen Position Recall
- [TEM Center: Upper menu] Control → Nanospace Map
- Display Nanospace Map
- Select the specimen position from the specimen position memory list
- Detailed information is displayed in the specimen position information display area
- Click [Go] button
- The specimen position is recalled.
- Wait until the recall is complete.
- Do not operate [TB] or [SC] during recall.
- [TEM Center: Upper menu] Control → Nanospace Map