Electron Beam Generation
- Preliminary Check
- Accelerating voltage / Step-up (if necessary) (Reference) Accelerating voltage setting
- Dark current check
💡 (Dark current) Beam Current is 100±3 (200kV)
- Holder insertion (Reference) Specimen setting
- [TEM Center: Upper menu] View → Valve/Vacuum Monitor
- Display Valve/Vacuum Monitor
- Check the vacuum level
- Penning Gauge 20 or less
- SIP 10^-5Pa range
💡 Do not generate an electron beam if the vacuum is poor The filament life may be shortened.
- [TEM Center: Upper menu] Control → Beam
- Display Beam Controller (for JEM Administrators)
💡 If the Filament:[ON] button is grayed out, contact the administrator.
- Electron beam generation
- Display Beam Controller (for JEM Administrators)
- Electron beam confirmation
- Confirm that the electron beam is displayed on the screen.
- If the electron beam is not displayed
💡 If the electron beam is not displayed even after doing the above, contact the administrator (Perform beam centering and alignment)
- Confirm that the electron beam is displayed on the screen.
Electron Beam Stopping
- [TEM Center: Upper menu] Control → Beam
- Display Beam Controller (for JEM Administrators)
- Electron beam stopping
- Holder removal (if necessary)
- Accelerating voltage / Standby (if necessary)
- Set to standby voltage (160kV) with Beam Controller / HT Voltage.